An excimer laser is a gas laser that emits ultraviolet pulses from an electrical discharge in a rare-gas and halogen mixture. Molecules that exist only in the excited state (excimers) emit light as they dissociate, and the gas pair sets the wavelength.

GasWavelengthTypical uses
ArF193 nmCorneal surgery, semiconductor lithography, micromachining of glass and fluoropolymers, analytical sampling
KrF248 nmFibre Bragg grating (FBG) writing, polymer micromachining, marking, pulsed laser deposition
XeCl308 nmLow-temperature polysilicon annealing, dermatology, surface treatment
XeF351 nmMicromachining, research

Why ultraviolet. The shorter the wavelength, the higher the energy of each photon. A 193 nm photon carries enough energy to break bonds in polymers and glass directly, so material is removed in thin layers from the surface without melting. Materials transparent to visible and near-infrared light still absorb at the surface and can be processed.

Pulse characteristics. Pulse widths are typically a few to a few tens of nanoseconds, pulse energy is high, and the beam has a wide rectangular cross-section. This suits projecting a mask pattern in one shot better than writing with a focused spot.

What to check when selecting

  • Wavelength: how strongly the material absorbs it comes first. Weak absorption gives rough edges and heat build-up.
  • Pulse energy and repetition rate: pulse energy matters for processing a large area per shot, repetition rate for throughput. Their product is the average power.
  • Energy stability: pulse-to-pulse variation changes the ablation depth. Check the stabilisation function and the sigma value.
  • Beam size and divergence: these go straight into the design of the mask projection optics.
  • Gas lifetime and maintenance: the halogen gas is consumed over time and must be exchanged. Check the gas exchange interval, tube lifetime and how windows are replaced.
  • Installation: power supply (single or three phase), cooling (air or water), size and weight, gas supply.

Compact excimer lasers. For OEM integration into equipment, smaller models that run on single-phase power are used instead of large industrial units. The MLase MLI series, for example, has 193 nm (ArF) and 248 nm (KrF) models with maximum repetition rates of 200, 500 or 1000 Hz, air or water cooling and 230 V single-phase power. Pulse width is 5 to 10 ns and energy stability is below 2% (sigma).

Shinhotek is the authorized distributor of MLase in Korea and supports specification review and supply of ArF and KrF excimer lasers.