A beam profiler measures the intensity distribution across a laser beam and reports beam size, centroid, ellipticity and uniformity. Four properties of the laser decide the choice: wavelength, beam size, power and pulse format.

Choose the measurement method first.

  • Camera-based (CCD or CMOS array): captures the whole cross-section at once. Best for 2D distribution, mode structure and uniformity, and the usual choice in industry.
  • Scanning (slit, knife-edge, pinhole): moves an aperture across a single detector and reconstructs the profile. Used for beams of a few micrometres or wavelengths a camera cannot sense.

Wavelength is set by the sensor material. Silicon sensors respond from roughly 350 to 1,100 nm. Near-infrared beyond 1,100 nm (for example 1,550 nm) needs InGaAs; ultraviolet or 10.6 µm CO2 beams need converter plates or thermal detectors. Outside the sensor range the sensitivity collapses and readings cannot be trusted.

Match beam size to pixel count. A beam larger than the active area is clipped; a beam that covers only a few pixels gives a coarse width. Check sensor size and pixel pitch so enough pixels span the beam, and add magnifying or reducing optics where needed.

Bring power down with attenuation. Camera sensors saturate around milliwatt levels, so ND filters, wedge samplers or beam splitters reduce the intensity. The attenuator must not alter the beam shape (fringes, thermal lensing). Pulsed lasers need a trigger input to synchronise exposure with the pulse.

ItemWhat to checkFailure mode
WavelengthSensor response rangeLow sensitivity, noise inflates the width
Beam sizeActive area, pixel pitchClipping or too few pixels
PowerSaturation limit, attenuatorSaturated, flattened profile
PulsesTrigger input, exposureMissed pulses, intensity jitter
StandardISO 11146 D4σ supportWidths differ between instruments

Agree on the beam-width definition. ISO 11146 uses the second-moment (D4σ) width, and M² needs widths at several positions around the focus. Know whether the software reports D4σ, 1/e² or FWHM before comparing instruments.

Shinhotek's LUMOS beam profilers are camera-based and the Lumosity software computes ISO 11146 widths and M². Tell us the wavelength, beam size and power and we will propose the sensor and attenuation setup.